suPARnostic® TurbiLatex
For rapid and high volume suPAR detectionAbout suPARnostic® TurbiLatex
CE-IVD marked, quantitative measurement of suPAR in human plasma, on the automated chemistry analyzers your lab already uses
suPARnostic® TurbiLatex is a fully automated suPAR assay for routine and high-volume testing. It runs in about 10 minutes on major Roche, Siemens, Abbott and Beckman Coulter platforms. There’s no separate instrument and no manual workflow: just a ready-to-use, two-reagent kit with dedicated calibrators and controls.
suPAR is a marker associated with immune activation and inflammation. Fast, automated results can support risk stratification and triage of patients in the emergency department and ICU.

Product details
A fully automated suPAR assay for routine clinical chemistry analyzers.
Assay
A particle-enhanced turbidimetric immunoassay (PETIA) for quantitative measurement of suPAR. Results in about 10 minutes, with no manual steps.
Sample type
Human K2-EDTA or lithium heparin plasma. The assay is calibrated for K2-EDTA. For lithium heparin plasma, a correction formula is provided in the Instructions for Use.
Measuring range
1.8–16.0 ng/mL, with demonstrated linearity up to 26.5 ng/mL. No prozone (hook) effect up to approximately 47.5 ng/mL, depending on the analyzer.
Sample stability
Plasma is stable for 24 hours at room temperature (20–25 °C) and 3 days at 2–8 °C. Freeze samples (≤ –20 °C) for longer storage.
Format & storage
Ready-to-use two-reagent kit. Store at 2–8 °C. Shelf life is 24 months, with onboard stability of about 8 weeks.
If you have any questions, we are happy to answer!
How it works
A latex particle-enhanced turbidimetric immunoassay in three steps.
1. Mix
The analyzer mixes the plasma sample with Reagent 1 (dilution buffer), then adds Reagent 2: latex particles coated with anti-suPAR antibodies.
2. Bind
suPAR in the sample binds to the antibody-coated particles, causing them to clump together. The more suPAR in the sample, the more clumping, and the cloudier (more turbid) the sample becomes.
3. Read & calculate
The analyzer measures the change in turbidity photometrically and calculates the suPAR concentration (ng/mL) against a calibration curve set up with the suPARnostic® TurbiLatex Calibrators. Total runtime is about 10 minutes.
It works like the other routine turbidimetric tests your lab already runs, so it fits straight into existing workflows.
If you have any questions, we are happy to answer!
Why choose suPARnostic® TurbiLatex
Built for fast, routine suPAR testing at scale.
Fast and automated
Results in about 10 minutes, with no manual pipetting, incubation or plate handling.
Runs on your existing analyzers
Validated on major Roche, Siemens, Abbott and Beckman Coulter platforms. No new instrument needed.
Standard sample tubes
Works with K2-EDTA and lithium heparin plasma, the tube types already used in routine blood sampling.
Reliable performance
Linear up to 26.5 ng/mL, with no prozone effect at clinically relevant concentrations, so high-suPAR samples are measured accurately.
Comparable with suPARnostic® ELISA
Results correlate closely with the suPARnostic® ELISA method (correlation coefficient ~0.96 in comparative studies).
Simple quality control
Dedicated calibrators and three-level controls. Calibrate at least monthly and whenever you start a new reagent lot.
If you have any questions, we are happy to answer!
What’s in the kit
suPARnostic® TurbiLatex Reagents (T001)
- Reagent 1: Dilution buffer
- Reagent 2: Latex particles coated with anti-suPAR antibodies
suPARnostic® TurbiLatex Calibrators (T020)
- Six ready-to-use liquid calibrators, 1 mL each
- Stored frozen (–18 to –20 °C), shelf life 12 months
suPARnostic® TurbiLatex Controls (T030)
- Three ready-to-use controls (low, medium, high), 1 mL each, with certificate of analysis
- Stored frozen (–18 to –20 °C), shelf life 12 months
Instructions for use and certificates are in the → technical documents
If you have any questions, we are happy to answer!
Validated platforms
suPARnostic® TurbiLatex is validated on the following clinical chemistry systems:
- Roche: cobas c 111, c 303, c 501/502, c 701/702, cobas pure, cobas pro
- Siemens: Atellica, ADVIA Chemistry XPT
- Abbott: ARCHITECT c, Alinity ci
- Beckman Coulter: DxC 5800 (verify the model name against the IFU)
Technical documents
Learn more about instructions for use and certificates from our documentation. Below you can see user instructions, quick guides, and links for software updates. If you have more questions, check out our FAQ section or do not hesitate to contact us!